Skip to main navigation Skip to search Skip to main content

World-to-chip microfluidic interface with built-in valves for multichamber chip-based PCR assays

  • Kwang W. Oh
  • , Chinsung Park
  • , Kak Namkoong
  • , Jintae Kim
  • , Kyeong Sik Ock
  • , Suhyeon Kim
  • , Young A. Kim
  • , Yoon Kyoung Cho
  • , Christopher Ko
  • Samsung

Research output: Contribution to journalArticlepeer-review

58 Scopus citations

Abstract

We report a practical world-to-chip microfluidic interfacing method with built-in valves suitable for microscale multichamber chip-based assays. One of the primary challenges associated with the successful commercialization of fully integrated microfluidic systems has been the lack of reliable world-to-chip microfluidic interconnections. After sample loading and sealing, leakage tests were conducted at 100 °C for 30 min and no detectable leakage flows were found during the test for 100 microchambers. To demonstrate the utility of our world-to-chip microfluidic interface, we designed a microscale PCR chip with four chambers and performed PCR assays. The PCR results yielded a 100% success rate with no contamination or leakage failures. In conclusion, we have introduced a simple and inexpensive microfluidic interfacing system for both sample loading and sealing with no dead volume, no leakage flow and biochemical compatibility.

Original languageEnglish
Pages (from-to)845-850
Number of pages6
JournalLab on a Chip
Volume5
Issue number8
DOIs
StatePublished - Aug 2005

Fingerprint

Dive into the research topics of 'World-to-chip microfluidic interface with built-in valves for multichamber chip-based PCR assays'. Together they form a unique fingerprint.

Cite this