Skip to main navigation Skip to search Skip to main content

Spatially resolved measurements to improve analytical performance of solution-cathode glow discharge optical-emission spectrometry

  • Andrew J. Schwartz
  • , Steven J. Ray
  • , George C.Y. Chan
  • , Gary M. Hieftje
  • Indiana University Bloomington
  • Lawrence Berkeley National Laboratory

Research output: Contribution to journalArticlepeer-review

32 Scopus citations

Abstract

Past studies of the solution-cathode glow discharge (SCGD) revealed that elemental and molecular emission are not spatially homogenous throughout the source, but rather conform to specific zones within the discharge. Exploiting this inhomogeneity can lead to improved analytical performance if emission is collected only from regions of the discharge where analyte species emit strongly and background emission (from continuum, elemental and/or molecular sources) is lower. Effects of this form of spatial discrimination on the analytical performance of SCGD optical emission spectrometry (OES) have been investigated with an imaging spectrograph for fourteen atomic lines, with emphasis on detection limits and precision. Vertical profiles of the emission intensity, signal-to-background ratio, and signal-to-noise ratio were collected and used to determine the optimal region to view the SCGD on a per-element basis. With optimized spatial filtering, detection limits ranged from 0.09–360 ppb, a 1.4–13.6 fold improvement over those obtained when emission is collected from the full vertical profile (1.1–840 ppb), with a 4.2-fold average improvement. Precision was found to be unaffected by spatial filtering, ranging from 0.5–2.6% relative standard deviation (RSD) for all elements investigated, closely comparable to the 0.4–2.4% RSD observed when no spatial filtering is used. Spatial profiles also appear useful for identifying optimal line pairs for internal standardization and for flagging the presence of matrix interferences in SCGD-OES.

Original languageEnglish
Pages (from-to)168-176
Number of pages9
JournalSpectrochimica Acta - Part B Atomic Spectroscopy
Volume125
DOIs
StatePublished - Nov 1 2016

Keywords

  • Instrumentation
  • Optical-emission spectrometry
  • Solution-cathode glow discharge
  • Spatial discrimination

Fingerprint

Dive into the research topics of 'Spatially resolved measurements to improve analytical performance of solution-cathode glow discharge optical-emission spectrometry'. Together they form a unique fingerprint.

Cite this