Skip to main navigation Skip to search Skip to main content

Safety and efficacy of the Pipeline embolization device for treatment of intracranial aneurysms: A pooled analysis of 3 large studies

  • David F. Kallmes
  • , Waleed Brinjikji
  • , Saruhan Cekirge
  • , David Fiorella
  • , Ricardo A. Hanel
  • , Pascal Jabbour
  • , Demetrius Lopes
  • , Pedro Lylyk
  • , Cameron G. McDougall
  • , Adnan Siddiqui
  • Mayo Clinic Rochester, MN
  • Stony Brook University
  • SUNY Buffalo
  • Lyerly Neurosurgery
  • Thomas Jefferson University
  • Rush University Medical Center
  • St. Joseph's Hospital and Medical Center, Phoenix
  • Koru Health Group

Research output: Contribution to journalArticlepeer-review

251 Scopus citations

Abstract

OBJECTIVE The authors performed a pooled analysis of 3 studies-IntrePED (International Retrospective Study of the Pipeline Embolization Device), PUFS (Pipeline for Uncoilable or Failed Aneurysms Study), and ASPIRe (Aneurysm Study of Pipeline in an Observational Registry)-in order to assess angiographic outcomes and clinical safety of the Pipeline embolization device (PED). METHODS IntrePED was a retrospective study, while PUFS and ASPIRe were prospective studies. For each patient included in these studies, the authors collected baseline demographic data, aneurysm characteristics, and procedural details. The primary outcomes for this combined analysis were clinical outcomes, including neurological morbidity and mortality and major ipsilateral intracranial hemorrhage and ischemic stroke. The secondary outcomes were angiographic occlusion rates, which were available for ASPIRe and PUFS only. RESULTS A total of 1092 patients with 1221 aneurysms were included across the 3 studies. The mean aneurysm size was 12.0 ± 7.8 mm and the mean neck size was 6.6 ± 4.8 mm. The major ipsilateral ischemic stroke rate was 3.7% (40/1091). The major ipsilateral intracranial hemorrhage rate was 2.0% (22/1091). The major neurological morbidity rate was 5.7% (62/1091). The neurological mortality rate was 3.3% (36/1091). The combined major morbidity and neurological mortality rate was 7.1% (78/1091). The complete occlusion rates were 75.0% at 180 days (111/148) and 85.5% at 1 year (94/110). The overall aneurysm retreatment rate was 3.0% (33/1091) at a mean follow-up time of 10.2 ± 10.8 months. CONCLUSIONS Endovascular treatment of intracranial aneurysms with the PED is safe and effective. Angiographic occlusion rates progressed with follow-up. Rates of stroke, hemorrhage, morbidity and mortality, and retreatment were low, especially given the fact that the aneurysms treated were generally large and wide necked.

Original languageEnglish
Pages (from-to)775-780
Number of pages6
JournalJournal of Neurosurgery
Volume127
Issue number4
DOIs
StatePublished - Oct 2017

Keywords

  • Aneurysm
  • Endovascular
  • Interventional neurosurgery
  • Stroke
  • Vascular disorders

Fingerprint

Dive into the research topics of 'Safety and efficacy of the Pipeline embolization device for treatment of intracranial aneurysms: A pooled analysis of 3 large studies'. Together they form a unique fingerprint.

Cite this