Skip to main navigation Skip to search Skip to main content

Reversed-phase high-performance liquid chromatograpic determination of verofylline in rat serum

  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

Abstract

A simple and selective high-performance liquid chromatographic procedure has been developed for the determination of verofylline in rat serum. This assay procedure, which is suitable for the analysis of microsamples (0.05 or 0.1 ml serum), involves a single extraction followed by isocratic chromatography on an RP-18 analytical column with ultraviolet detection at 280 nm. The calibration curve is linear within the study range 0.25-75 μg/ml. The intra-day and inter-day coefficients of variation were below 5.7%. The specificity of the assay was confirmed by comparing the lipophilicity, absorbance, and chromatographic property of the apparent and authentic verofylline.

Original languageEnglish
Pages (from-to)207-213
Number of pages7
JournalJournal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences
VolumeVol. 382
StatePublished - 1986

Fingerprint

Dive into the research topics of 'Reversed-phase high-performance liquid chromatograpic determination of verofylline in rat serum'. Together they form a unique fingerprint.