Skip to main navigation Skip to search Skip to main content

Plenoptic cameras in surgical robotics: Calibration, registration, and evaluation

  • Azad Shademan
  • , Ryan S. Decker
  • , Justin Opfermann
  • , Simon Leonard
  • , Peter C.W. Kim
  • , Axel Krieger

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

36 Scopus citations

Abstract

Three-dimensional sensing of changing surgical scenes would improve the function of surgical robots. This paper explores the requirements and utility of a new type of depth sensor, the plenoptic camera, for surgical robots. We present a metric calibration procedure for the plenoptic camera and the registration of its coordinate frame to the robot (hand-eye calibration). We also demonstrate the utility in robotic needle insertion and application of sutures in phantoms. The metric calibration accuracy is reported as 1.14 ± 0.80 mm for the plenoptic camera and 1.57 ± 0.90 mm for hand-eye calibration. The accuracy of needle insertion task is 1.79 ± 0.35 mm for the entire robotic system. Additionally, the accuracy of suture placement with the presented system is reported at 1.80 ± 0.43 mm. Finally, we report consistent suture spacing with only 0.11 mm standard deviation between inter-suture distances. The measured accuracy of less than 2 mm with consistent suture spacing is a promising result to provide repeatable leak-free suturing with a robotic tool and a plenoptic depth imager.

Original languageEnglish
Title of host publication2016 IEEE International Conference on Robotics and Automation, ICRA 2016
PublisherInstitute of Electrical and Electronics Engineers Inc.
Pages708-714
Number of pages7
ISBN (Electronic)9781467380263
DOIs
StatePublished - Jun 8 2016
Event2016 IEEE International Conference on Robotics and Automation, ICRA 2016 - Stockholm, Sweden
Duration: May 16 2016May 21 2016

Publication series

NameProceedings - IEEE International Conference on Robotics and Automation
Volume2016-June
ISSN (Print)1050-4729

Conference

Conference2016 IEEE International Conference on Robotics and Automation, ICRA 2016
Country/TerritorySweden
CityStockholm
Period05/16/1605/21/16

Fingerprint

Dive into the research topics of 'Plenoptic cameras in surgical robotics: Calibration, registration, and evaluation'. Together they form a unique fingerprint.

Cite this