Abstract
The accuracy that can be achieved in single-pulse pump-probe Laue experiments is discussed. It is shown that with careful tuning of the experimental conditions a reproducibility of the intensity ratios of equivalent intensities obtained in different measurements of 3-4% can be achieved. The single-pulse experiments maximize the time resolution that can be achieved and, unlike stroboscopic techniques in which the pump-probe cycle is rapidly repeated, minimize the temperature increase due to the laser exposure of the sample.
| Original language | English |
|---|---|
| Pages (from-to) | 479-485 |
| Number of pages | 7 |
| Journal | Journal of Synchrotron Radiation |
| Volume | 17 |
| Issue number | 4 |
| DOIs | |