Skip to main navigation Skip to search Skip to main content

Novel Flow-Through Pneumatoamperometric Detector for Determination of Nanogram and Subnanogram Amounts of Nitrite by Flow-Injection Analysis

  • Czechoslovak Acadamy of Sciences

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

A gas porous electrode structure that detects volatile electroactive species in a flowing liquid stream is described and evaluated for its utility in flow injection analysis. The electrode is fabricated by depositing a porous gold layer on one side of a porous Teflon membrane. The gold serves as the amperometric electrode which consumes dissolved, volatile species that is transported from the flowing solution through the membrane to the metallzed face where it is electrolyzed. Nitrite ion Is determined by reaction In the carrier stream to produce nitric oxide and iodine, and both are electroxidized at the gold electrode. The detection limit is 30 pg of nitrite ion. Dissolved, nonvolatile electroactive species do not Interfere.

Original languageEnglish
Pages (from-to)866-869
Number of pages4
JournalAnalytical Chemistry
Volume58
Issue number4
DOIs
StatePublished - Apr 1986

Fingerprint

Dive into the research topics of 'Novel Flow-Through Pneumatoamperometric Detector for Determination of Nanogram and Subnanogram Amounts of Nitrite by Flow-Injection Analysis'. Together they form a unique fingerprint.