NEW TECHNIQUE FOR THE ELIMINATION OF FLUORESCENCE INTERFERENCE IN RAMAN SPECTROSCOPY.

  • Indiana University Bloomington

Research output: Contribution to journalArticlepeer-review

20 Scopus citations

Abstract

A new instrument for Raman spectroscopy has been constructed with the use of a cw or mode-locked argon-ion laser as the excitation source and a microwave spectrum analyzer for discrimination between fluorescence and Raman scatter. To demonstrate the new instrument's capability, the weak Raman scatter of water was resolved from the intense fluorescence from rhodamine 6G. Theoretically predicted relative errors in the new procedure are calculated and experimental results are shown to be in good agreement with theory.

Original languageEnglish
Pages (from-to)583-587
Number of pages5
JournalApplied Spectroscopy
Volume40
Issue number5
DOIs
StatePublished - 1986