Abstract
We developed a microfluidic chip that provides rapid temperature changes and accurate temperature control of the perfusing solution to facilitate patch-clamp studies. The device consists of a fluid channel connected to an accessible reservoir for cell culture and patch-clamp measurements. A thin-film platinum heater was placed in the flow channel to generate rapid temperature change, and the temperature was monitored using a thin-film resistor. We constructed the thermal chip using SU-8 on a glass wafer to minimize the heat loss. The chip is capable of increasing the solution temperature from bath temperature (20°C) to 80°C at an optimum heating rate of 0.5°C/ms. To demonstrate the ability of the thermal chip, we have conducted on-chip patch-clamp recordings of temperature-sensitive ion channels (TRPV1) transfected HEK293 cells. The heat-stimulated currents were observed using whole-cell and cell-attached patch configurations. The results demonstrated that the chip can provide rapid temperature jumps at the resolution of single-ion channels.
| Original language | English |
|---|---|
| Pages (from-to) | 2447-2451 |
| Number of pages | 5 |
| Journal | Analytical Chemistry |
| Volume | 80 |
| Issue number | 7 |
| DOIs | |
| State | Published - Apr 1 2008 |
Fingerprint
Dive into the research topics of 'Microfluidic chip to produce temperature jumps for electrophysiology'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver