Measurement of Flow Modification in Phantom Aneurysm Model: Comparison of Coils and a Longitudinally and Axially Asymmetric Stent - Initial Findings

  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

45 Scopus citations

Abstract

Dye-dilution imaging sequences were performed and time-density curves were constructed in elastomer vessel aneurysm models to demonstrate the effectiveness of coils and an asymmetric stent in disrupting standard vortex flow. Compared with the use of coils, the use of stents led to marked flow modification, as seen with imaging sequences, and substantially slower inflow, as indicated by time-density curves, owing to the low-porosity region of the stent that covers the aneurysm orifice. These flow examination results indicate that potentially favorable flow modification features can be created by using the described asymmetric stent design, the use of which may lead to alternative methods of image-guided endovascular cerebral aneurysm therapy.

Original languageEnglish
Pages (from-to)272-276
Number of pages5
JournalRadiology
Volume231
Issue number1
DOIs
StatePublished - Apr 2004

Keywords

  • Aneurysm, cerebral
  • Aneurysm, therapy
  • Angiography
  • Blood
  • Experimental study
  • Flow dynamics
  • Phantoms
  • Stents and prostheses