Skip to main navigation Skip to search Skip to main content

Ion cyclotron resonance bridge detector for frequency sweep

  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

An electronic ion cyclotron resonance detection system was designed and constructed. The ions are excited by sweeping the frequency of the electric field (3-300 kHz) using a sweep frequency generator with a nonlinear sweep voltage in order to maintain an approximately constant mass resolution. Ion detection is accomplished by a bridge with a phase-sensitive detector as a demodulator. The required reference signal for the phase-sensitive detector is generated by a circuit with a transfer function which approximates that of the ICR signal in order to obtain an accurate phase match between the signal source and the detector. The device is capable of detecting a minimum concentration of 50 ions/cm3 over a mass range of 15 to 1500 amu.

Original languageEnglish
Pages (from-to)1476-1481
Number of pages6
JournalReview of Scientific Instruments
Volume54
Issue number11
DOIs
StatePublished - 1983

Fingerprint

Dive into the research topics of 'Ion cyclotron resonance bridge detector for frequency sweep'. Together they form a unique fingerprint.

Cite this