Abstract
A consecutive series of twenty-seven patients who had thirty acute unstable (type-C) fractures of the pelvic ring was studied prospectively to evaluate the use of stimulus-evoked electromyography to decrease the risk of iatrogenic nerve-root injury during the insertion of iliosacral screws. A prerequisite for inclusion in the study was a normal neurological status preoperatively; somatosensory evoked potentials were monitored to further document the neurological status both before and after insertion of the screw or screws. A total of fifty-one iliosacral screws were inserted, and a current threshold of more than eight milliamperes was selected as the level that indicated that the drill-bit was a safe distance from the nerve root. Four of the fifty-one screws were redirected because of information obtained with stimulus-evoked electromyography. Postoperatively, all patients had a normal neurological status. Computerized tomography, although not accurate for detailed measurements, demonstrated that all of the screws were in a safe, intraosseous position. Monitoring with stimulus-evoked electromyography appears to provide reliable data and may decrease the risk of iatrogenic injury to the nerve roots during operations on the pelvic ring.
| Original language | English |
|---|---|
| Pages (from-to) | 537-546 |
| Number of pages | 10 |
| Journal | Journal of Bone and Joint Surgery |
| Volume | 80 |
| Issue number | 4 |
| DOIs | |
| State | Published - Apr 1998 |
Fingerprint
Dive into the research topics of 'Intraoperative monitoring with stimulus-evoked electromyography during placement of iliosacral screws: An initial clinical study'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver