Abstract
Introduction: The purpose of this study was to report our initial experience of using a large-bore (0.096″ inner diameter) access catheter in neurovascular interventions. Methods: Data were retrospectively collected from 5 sites in the US for neurovascular procedures performed using a large-bore access catheter. The effectiveness outcome was technical success, defined as the access catheter's successfully reaching its target vessel without conversion to direct carotid puncture or to a smaller-bore access catheter and successfully completing the intended neurointervention. Results: One hundred and thirteen procedures performed in 112 patients were included in this study. The mean age of the patients was 67.5 years (SD 16.2), and about half (49.1%) were female. The most common primary access sites were the femoral (64.6%) or radial (32.7%) artery. Challenging anatomic variations included severe vessel tortuosity (26/81, 32.1%), type II aortic arch (17/88, 19.3%), type III aortic arch (14/88, 15.9%), bovine arch (16/104, 15.4%), severe angle (<30°) between the subclavian and target vessel (11/74, 14.9%), and subclavian loop (7/79, 8.9%). The median access time to branch view was 18 min (IQR 11-28, N = 75). The technical success rate was 94.7%. Two dissections (1.8%) were related to the large-bore access catheter. Access site complications occurred in 2 patients (1.8%). Four additional symptomatic periprocedural complications not related to the large-bore access catheter occurred (7.1%). Conclusion: For neurovascular interventions, a 0.096″ inner diameter access catheter could be used with both femoral and radial arterial approaches, had a high technical success rate, and had a low rate of periprocedural complications.
| Original language | English |
|---|---|
| Pages (from-to) | 372-379 |
| Number of pages | 8 |
| Journal | Interventional Neuroradiology |
| Volume | 30 |
| Issue number | 3 |
| DOIs | |
| State | Published - Jun 2024 |
Keywords
- Access catheter
- catheters
- coiling
- endovascular procedures
- guide catheter
- interventional radiology
- ischemic stroke
- neurovascular interventions
- thrombectomy
Fingerprint
Dive into the research topics of 'Initial experience of using a large-bore (0.096″ inner diameter) access catheter in neurovascular interventions'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver