Abstract
A high‐pressure liquid chromatographic method for the sensitive, rapid, and specific determination of sulfapyridine and its N‐acetyl derivative in plasma and saliva was developed. A cyano‐bonded, reversed‐phase, high efficiency column was used. The system detected these sulfonamides in serum to 0.25 mg/liter and within only 6 min. Sulfapyridine was separated from its acetyl derivative with little interference from other drugs. The assay reproducibility was with 3%. The assay was highly useful for routine monitoring of patients receiving sulfasalazine for inflammatory bowel disease.
| Original language | English |
|---|---|
| Pages (from-to) | 1250-1253 |
| Number of pages | 4 |
| Journal | Journal of Pharmaceutical Sciences |
| Volume | 67 |
| Issue number | 9 |
| DOIs | |
| State | Published - Sep 1978 |
Keywords
- Dermatitis suppressants—sulfapyridine and acetyl metabolite, high‐performance liquid chromatographic analyses in biological fluids
- High‐performance liquid. chromatography—analyses, sulfapyridine and acetyl metabolite in biological fluids
- Sulfapyridine and acetyl metabolite—high‐performance liquid chromatographic analyses in biological fluids
Fingerprint
Dive into the research topics of 'High‐performance liquid chromatographic assay of sulfapyridine and acetylsulfapyridine in biological fluids'. Together they form a unique fingerprint.Cite this