Skip to main navigation Skip to search Skip to main content

High-performance liquid chromatographic assay for the determination of total and free topotecan in the presence and absence of anti-topotecan antibodies in mouse plasma

  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

23 Scopus citations

Abstract

A rapid and sensitive high-performance liquid chromatographic (HPLC) assay has been developed to allow determination of total (i.e. bound and unbound) and free (i.e. unbound) topotecan (TPT) in mouse plasma in the presence and absence of anti-TPT antibodies. The chromatographic analysis was carried out using reversed-phase isocratic elution with a Nova-Pak C18 column (3.9 mm × 150 mm, 4 μm) protected by a Nova-Pak C18 guard column (3.9 mm × 20 mm, 4 μm), where 10 mM KH2PO4-methanol-triethylamine (72:26:2 (v/v/v), pH 3.5) was used as the mobile phase. Topotecan was quantified with fluorescence detection using an excitation wavelength of 361 nm and an emission wavelength of 527 nm. The retention time for the internal standard, acridine, and TPT were 7.4 and 9.0 min, respectively. The lower limit of quantitation (LOQ) for TPT was determined as 0.02 ng in mouse plasma and mouse plasma ultrafiltrate, corresponding to a concentration of 1 ng/ml in 20 μl mouse plasma. The assay was shown to be linear over a concentration range of 1-500 ng/ml. The recoveries of free and total TPT from spiked mouse plasma were within 10% of theoretical values (assessed at 1, 20 and 500 ng/ml). The validated HPLC assay was applied to evaluate TPT pharmacokinetics following administration of TPT to Swiss Webster mice and to hyperimmunized and control BALB/c mice. The assay has been shown to be capable for measuring total and free TPT in mouse plasma with high sensitivity and will allow the testing of the effect of anti-TPT antibodies on the disposition of TPT.

Original languageEnglish
Pages (from-to)183-192
Number of pages10
JournalJournal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences
Volume816
Issue number1-2
DOIs
StatePublished - Feb 25 2005

Keywords

  • Anti-TPT antibodies
  • HPLC
  • Topotecan

Fingerprint

Dive into the research topics of 'High-performance liquid chromatographic assay for the determination of total and free topotecan in the presence and absence of anti-topotecan antibodies in mouse plasma'. Together they form a unique fingerprint.

Cite this