Skip to main navigation Skip to search Skip to main content

Graphite-halogens as temperature calibration standards for transmission electron microscopy

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Graphite-halogen (Br2, ICl) residue compounds have been found to be convenient temperature calibration standards for transmission electron microscopy in the medium temperature range. Due to the order-disorder transformation associated with the intralayer intercalate ordering in these intercalation compounds, the observation of the disappearance of the superlattice diffraction spots as the sample temperature is raised serves to calibrate the temperature. The transformation occurs at 100±1°C in graphite-Br2 and 41±1°C in graphite-ICl. The simplicity of specimen preparation gives added attraction to the use of graphite-halogens as temperature calibration standards.

Original languageEnglish
Pages (from-to)932-934
Number of pages3
JournalReview of Scientific Instruments
Volume51
Issue number7
DOIs
StatePublished - 1980

Fingerprint

Dive into the research topics of 'Graphite-halogens as temperature calibration standards for transmission electron microscopy'. Together they form a unique fingerprint.

Cite this