title = "Focal spot size reduction using asymmetric collimation to enable reduced anode angles with a conventional angiographic x-ray tube for use with high resolution detectors",
abstract = "The high-resolution requirements for neuro-endovascular image-guided interventions (EIGIs) necessitate the use of a small focal-spot size; however, the maximum tube output limits for such small focal-spot sizes may not enable sufficient x-ray fluence after attenuation through the human head to support the desired image quality. This may necessitate the use of a larger focal spot, thus contributing to the overall reduction in resolution. A method for creating a higher-output small effective focal spot based on the line-focus principle has been demonstrated and characterized. By tilting the C-arm gantry, the anode-side of the x-ray field-of-view is accessible using a detector placed off-axis. This tilted central axis diminishes the resultant focal spot size in the anode-cathode direction by the tangent of the effective anode angle, allowing a medium focal spot to be used in place of a small focal spot with minimal losses in resolution but with increased tube output. Images were acquired of two different objects at the central axis, and with the C-arm tilted away from the central axis at 1° increments from 0°-7°. With standard collimation settings, only 6° was accessible, but using asymmetric extended collimation a maximum of 7° was accessed for enhanced comparisons. All objects were positioned perpendicular to the anode-cathode direction and images were compared qualitatively. The increasing advantage of the off-axis focal spots was quantitatively evidenced at each subsequent angle using the Generalized Measured-Relative Object Detectability metric (GM-ROD). This anode-tilt method is a simple and robust way of increasing tube output for a small field-of-view detector without diminishing the overall apparent resolution for neuro-EIGIs.",
keywords = "Generalized Measured-Relative Object De- tectability (GM-ROD), High Resolution Fluoroscope (HRF), Line Focus Principle, Tilted Anode",
author = "M. Russ and A. Shankar and \{Setlur Nagesh\}, \{S. V.\} and Ionita, \{C. N.\} and Bednarek, \{D. R.\} and S. Rudin",
note = "Publisher Copyright: {\textcopyright} 2017 SPIE.; Medical Imaging 2017: Physics of Medical Imaging ; Conference date: 13-02-2017 Through 16-02-2017",
year = "2017",
doi = "10.1117/12.2253724",
language = "English",
series = "Progress in Biomedical Optics and Imaging - Proceedings of SPIE",
publisher = "SPIE",
editor = "Schmidt, \{Taly Gilat\} and Lo, \{Joseph Y.\} and Flohr, \{Thomas G.\}",
booktitle = "Medical Imaging 2017",
address = "United States",
}