Skip to main navigation Skip to search Skip to main content

Flow-pattern details in an aneurysm model using high-speed 1000-frames-per-second angiography

  • SUNY Buffalo
  • Women and Children's Hospital of Buffalo
  • Jacobs Institute
  • University of South Florida

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

SUMMARY: Traditional digital subtraction angiography provides rather limited evaluation of contrast flow dynamics when studying and treating intracranial brain aneurysms. A 1000-frames-per-second photon-counting x-ray detector was used to image detailed iodinecontrast flow patterns in an internal carotid artery aneurysm of a 3D-printed vascular phantom. High-speed imaging revealed differences in vortex and inflow patterns with and without a Pipeline Embolization Device flow diverter in more detail and clarity than could be seen in standard pulsed angiography. Improved temporal imaging has the potential to impact the outcomes of endovascular interventions by allowing clinicians to better understand and act on flow dynamics in real-time.

Original languageEnglish
Pages (from-to)1197-1209
Number of pages13
JournalAmerican Journal of Neuroradiology
Volume40
Issue number7
DOIs
StatePublished - Jul 1 2019

Fingerprint

Dive into the research topics of 'Flow-pattern details in an aneurysm model using high-speed 1000-frames-per-second angiography'. Together they form a unique fingerprint.

Cite this