Abstract
A laboratory test of X-ray tomography employing a diverging beam of X-rays rather than the usual parallel X-ray beam is described. We chose to test and demonstrate the advantages of divergent beam tomography by imaging an extracted juvenile human premolar using an ordinary dental X-ray source and a cooled CCD camera. Experiments with a three-piece cover-glass sample and with the human tooth demonstrated that three-dimensional reconstruction can be achieved at 34 μm per pixel resolution employing an X-ray tube spot 800 μm in its smallest direction without requiring close contact with the fluorescent screen. Reconstruction of a 256 x 256 pixel single-plane image from 100 projection images took only 45 sec on a personal computer with a Pentium 166 MHz processor. We have also demonstrated a volume reconstruction of 256 x 256 x 256 voxels from the data. Successful extension of this work to submicrometer projection X-ray microscopy is predicted. Improved resolution of medical tomography is another possible application.
| Original language | English |
|---|---|
| Pages (from-to) | 56-62 |
| Number of pages | 7 |
| Journal | Microscopy and Microanalysis |
| Volume | 4 |
| Issue number | 1 |
| DOIs | |
| State | Published - 1998 |
Keywords
- Cone-beam
- Microtomography
- Shadow projection X-ray microscopy
- Tomography
- X-ray