Abstract
The details of construction and operation of an alternating current plasma (ACP) detector supported with helium gas for reversed-phase high-performance liquid chromatography is described. The system is evaluated as a selective detector for the determination of organomercury compounds. The chromatographic eluent is introduced into the plasma by means of a frit nebulizer and the plasma can tolerate the nebulization of 100% methanol without extinguishment. Detection limits of methylmercury chloride and ethylmercury chloride were found to be 4.5 and 2.2 ng Hg/sec, respectively. In the nanogram to microgram mass range studied the precision was found to be less than 10% (relative standard deviation). The deduction of organomercurials in complex sample matrices elustrates the selectivity of the ACP detector.
| Original language | English |
|---|---|
| Pages (from-to) | 159-166 |
| Number of pages | 8 |
| Journal | Journal of Chromatography A |
| Volume | 513 |
| Issue number | C |
| DOIs | |
| State | Published - 1990 |
Fingerprint