Enhanced Mass Resolving Power, Sensitivity, and Selectivity in Laser Desorption Fourier Transform Ion Cyclotron Resonance Mass Spectrometry by Ion Axialization and Cooling

  • Shenheng Guan
  • , Markus C. Wahl
  • , Troy D. Wood
  • , Alan G. Marshall
  • Ohio State University
  • Department of Biochemistry

Research output: Contribution to journalArticlepeer-review

89 Scopus citations

Abstract

Ion cooling and axialization produced by azimuthal quadrupolar excitation in the presence of ion‒neutral collisions are applied to laser desorption Fourier transform ion cyclotron resonance mass spectrometry (LD/FT/ICR-MS). With this technique, the large kinetic and internal energies of ions generated by laser desorption processes can be cooled effectively by collisions of ions with neutral argon atoms (at > 5 × 10-7 Torr). After sufficient cooling in the source compartment of a dual ion trap, the axialized ions may be transferred to the analyzer compartment for detection at much lower pressure (and thus much higher mass resolving power). Enhancements in both FT/ICR mass resolving power and sensitivity are observed; moreover, ion isolation with high selectivity at high pressure is also demonstrated.

Original languageEnglish
Pages (from-to)1753-1757
Number of pages5
JournalAnalytical Chemistry
Volume65
Issue number13
DOIs
StatePublished - Jul 1 1993

Fingerprint

Dive into the research topics of 'Enhanced Mass Resolving Power, Sensitivity, and Selectivity in Laser Desorption Fourier Transform Ion Cyclotron Resonance Mass Spectrometry by Ion Axialization and Cooling'. Together they form a unique fingerprint.