Skip to main navigation Skip to search Skip to main content

Effects of Inaccurate Reference Lifetimes on Interpreting Frequency-Domain Fluorescence Data

  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

The effects of assigning Inaccurate reference lifetimes in lifetime determinations are predicted theoretically by using standard equations. This theory leads to a method to remove reference error effects using common least-squares software. This method cannot, however, be used to deconvoiute data collected with Isochronal references. Uncorrected data can always be exactly solved with models containing one more degree of freedom than the true model. Monoexponentlaf decays are fit by double-exponential decays or excited-state processes. Unlmodal distributed decays often appear as discrete, double-exponential decays.

Original languageEnglish
Pages (from-to)471-476
Number of pages6
JournalAnalytical Chemistry
Volume62
Issue number5
DOIs
StatePublished - Mar 1 1990

Fingerprint

Dive into the research topics of 'Effects of Inaccurate Reference Lifetimes on Interpreting Frequency-Domain Fluorescence Data'. Together they form a unique fingerprint.

Cite this