Dose reduction during fluoroscopic placement of feeding tubes

  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

By both increasing the optical iris of the video camera and removing the grid in fluoroscopic procedures involved in placement of a Dobbhoff feeding tube, the radiation dose to patients and staff was reduced by five to seven times. An average expected dose to a patient of about 300 mGy per procedure was reduced by two to three times when the grid was removed and by an additional three times when an iris of increased diameter was used. Because Dobbhoff procedures do not involve obtaining a diagnosis, the image degradation was acceptable and did not affect the total exposure times or ability to conduct the procedure. With the grid out, the difference between the mean exposure times of 5.1 minutes in 96 patient studies done with a normal iris opening and 4.0 minutes in 52 studies done with an enlarged iris was not statistically significant. The importance of reducing patient dose is reinforced by the finding that one-third of the patients underwent repeated procedures, accounting for almost 60% of the total.

Original languageEnglish
Pages (from-to)647-651
Number of pages5
JournalRadiology
Volume178
Issue number3
StatePublished - Mar 1991

Keywords

  • Diagnostic radiology, radiation exposure
  • Feeding tube
  • Radiations, exposure to patients personnel