Skip to main navigation Skip to search Skip to main content

Directly coupled direct current superconducting quantum interference device magnetometers based on ramp-edge Ag:YBa2Cu3O7-x/PrBa2Cu 3O7-x/Ag:YBa2Cu3O7-x junctions

  • Q. X. Jia
  • , F. Yan
  • , C. Mombourquette
  • , D. Reagor
  • Los Alamos National Laboratory

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

Directly coupled dc superconducting quantum interference device (SQUID) magnetometers on LaAlO3 substrates were fabricated using ramp-edge superconductor/normal-metal/superconductor junctions, where Ag-doped YBa2Cu3O7-x was used for the electrode and PrBa2Cu3O7-x for the normal-metal barrier. A flux noise of 8×10-60Hz-1/2 at 10 kHz measured with a dc bias current was achieved at 75 K, which corresponded to a field sensitivity of 400fTHz-1/2 for a magnetometer with a pick-up loop area of 8.5mm×7.5mm. Most significantly, the noise floor increased at lower frequencies with a frequency dependence slightly less than 1/f. The field noise of the SQUID magnetometers increased by only 25% after cycling the devices from zero field to 500 mG. In a static earth's magnetic field background, the field noise of the SQUID magnetometers increased by less than a factor of 2.

Original languageEnglish
Pages (from-to)3068-3070
Number of pages3
JournalApplied Physics Letters
Volume72
Issue number23
DOIs
StatePublished - 1998

Fingerprint

Dive into the research topics of 'Directly coupled direct current superconducting quantum interference device magnetometers based on ramp-edge Ag:YBa2Cu3O7-x/PrBa2Cu 3O7-x/Ag:YBa2Cu3O7-x junctions'. Together they form a unique fingerprint.

Cite this