Skip to main navigation Skip to search Skip to main content

Determination of adsorption-induced variation in the spring constant of a microcantilever

  • Oak Ridge National Laboratory

Research output: Contribution to journal › Article › peer-review

94 Scopus citations

Abstract

Although resonating microcantilevers are demonstrated to be excellent mass sensors, adsorption-induced changes in the spring constant result in errors in the calculation of adsorbed mass from shifts in resonance frequencies. However, simultaneous measurement of resonance frequency and adsorption-induced cantilever bending can be used to determine the variation in spring constant. Plotting the change in surface stress as a function of analyte concentration, the surface excess of adsorbed molecules and, therefore, the mass adsorbed can be determined. Here, we demonstrate this concept for adsorption of Na + ions on microcantilevers in NaCl solutions where a change in the spring constant was found to increase from 9.5×10-4 to 7.5×10-3 N/m as the NaCl concentration increased from 0.05 to 0.8 M.

Original languageEnglish
Pages (from-to)2219-2221
Number of pages3
JournalApplied Physics Letters
Volume80
Issue number12
DOIs
StatePublished - Mar 25 2002

Fingerprint

Dive into the research topics of 'Determination of adsorption-induced variation in the spring constant of a microcantilever'. Together they form a unique fingerprint.