Skip to main navigation Skip to search Skip to main content

Detection of HG2+ using microcantilever sensors

  • Louisiana Tech University
  • Oak Ridge National Laboratory

Research output: Contribution to journalArticlepeer-review

106 Scopus citations

Abstract

Trace amounts of Hg2+ are detected by using a microcan. tilever coated with gold. The microcantilever undergoes bending due to accumulation of Hg2+ on the gold surface. It is found that a concentration of 10-11 M Hg2+ can be detected using this technology. Other cations, such as K+, Na+, Pb2+, Zn2+, Ni2+, Cd2+, Cu2+, and Ca2+ have little or no effect on the deflection of the cantilever. The selectivity of the Hg2+ sensor could be improved by coating the gold surface of microcantilever with a self-assembled monolayer of a long-chain thiol compound.

Original languageEnglish
Pages (from-to)3611-3615
Number of pages5
JournalAnalytical Chemistry
Volume74
Issue number15
DOIs
StatePublished - Aug 1 2002

Fingerprint

Dive into the research topics of 'Detection of HG2+ using microcantilever sensors'. Together they form a unique fingerprint.

Cite this