Skip to main navigation Skip to search Skip to main content

Construction and performance of a simple multishielded Dewar

  • SUNY Buffalo

Research output: Contribution to journal › Article › peer-review

1 Scopus citations

Abstract

We have constructed a multishielded Dewar which we have tested at nitrogen and helium temperatures. The Dewar is of extremely simple design, requires no special insulating materials, and attains excellent performance. With an arrangement of 20 shields we have obtained evaporation rates of 5.6 and 3.0 liquid cm3 per hour for liquid helium and nitrogen, respectively. We report measurements of temperature distributions and evaporation rate as a function of the number of shields.

Original languageEnglish
Pages (from-to)236-238
Number of pages3
JournalReview of Scientific Instruments
Volume53
Issue number2
DOIs
StatePublished - 1982

Fingerprint