Abstract
Fundamental aspects of constant-momentum acceleration time-of-flight mass spectrometry (CMA-TOFMS) are explored as a means to improve mass resolution. By accelerating all ions to the same momentum rather than to the same energy, the effects of the initial ion spatial and energy distributions upon the total ion flight time are decoupled. This decoupling permits the initial spatial distribution of ions in the acceleration region to be optimized independently, and energy focus, including ion turn-around-time error, to be accomplished with a linear-field reflectron. Constant-momentum acceleration also linearly disperses ions across time according to mass-to-charge (m/z) ratio, instead of the quadratic relationship between flight time and m/z found in conventional TOFMS. Here, CMA-TOFMS is shown to achieve simultaneous spatial and energy focusing over a selected portion of the mass spectrum. An orthogonal- acceleration time-of-flight system outfitted with a reduced-pressure DC glow discharge (GD) ionization source is used to demonstrate CMA-TOFMS with atomic ions. The influence of experimental parameters such as the amplitude and width of the time-dependent CMA pulse on mass resolution is investigated, and a useful CMA-TOFMS focusing window of 2 to 18 Da is found for GD-CMA-TOFMS. [Figure not available: see fulltext.]
| Original language | English |
|---|---|
| Pages (from-to) | 1853-1861 |
| Number of pages | 9 |
| Journal | Journal of the American Society for Mass Spectrometry |
| Volume | 24 |
| Issue number | 12 |
| DOIs | |
| State | Published - Dec 2013 |
Keywords
- Constant-momentum acceleration
- Glow Discharge
- Mass spectrometry
- Time-of-flight
- Turn-around time
Fingerprint
Dive into the research topics of 'Constant-momentum acceleration time-of-flight mass spectrometry with energy focusing'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver