Comparison of the patient’s skin dose for flat and curved surfaces as a function of x-ray beam angle of incidence

  • SUNY Buffalo

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

It is important to determine the patient’s skin dose accurately for fluoroscopic interventional procedures in order to estimate the risk of deterministic injury. The purpose of this study is to investigate how the patient’s skin dose changes as a function of x-ray beam incident angle for flat and curved surfaces. The primary and scatter dose was calculated averaged over a 2.0 mm depth at the surfaces of both cubic and cylindrical water phantoms to simulate different patient curvature. The total skin dose was calculated using EGSnrc Monte-Carlo (MC) software with 1010 photons incident and the primary dose was calculated at the central axis using the mass energy absorption coefficients published by NIST and integrated over the beam-energy spectrum. Simulations were done for incident angles from 90 to 10 degrees, beam field sizes from 5 to 15 cm, cylinder diameters from 20 to 30 cm, and beam energies from 60 to 120 kVp. The results show the scatter-plus-primary to incident-primary dose ratio decreases with decreasing incident angle due to increased primary attenuation and decreases from cubic to cylindrical phantom and with decreasing cylinder diameter at all angles due to reduced backscatter. These results can be used to determine angular correction factors needed to accurately estimate patient skin dose when the beam is not normal to the entrance surface during fluoroscopic procedures.

Original languageEnglish
Title of host publicationMedical Imaging 2020
Subtitle of host publicationPhysics of Medical Imaging
EditorsGuang-Hong Chen, Hilde Bosmans
PublisherSPIE
ISBN (Electronic)9781510633919
DOIs
StatePublished - 2020
EventMedical Imaging 2020: Physics of Medical Imaging - Houston, United States
Duration: Feb 16 2020Feb 19 2020

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume11312
ISSN (Print)1605-7422

Conference

ConferenceMedical Imaging 2020: Physics of Medical Imaging
Country/TerritoryUnited States
CityHouston
Period02/16/2002/19/20

Keywords

  • Backscatter
  • Diagnostic x-ray dosimetry
  • Fluoroscopy
  • Interventional radiology
  • Monte Carlo simulation
  • Patient dose
  • Radiation dose
  • Skin dose
  • X-ray