Skip to main navigation Skip to search Skip to main content

Capillary electrochromatography in columns packed by gravity. Preliminary study

  • Kimberly J. Reynolds
  • , Todd D. Maloney
  • , Adam M. Fermier
  • , Luis A. Colón
  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

31 Scopus citations

Abstract

Chromatographic material was delivered into fused silica capillary columns mainly by the use of sedimentation by gravity. After sedimentation of the packing material, the columns were pressurized at 8000 psi and evaluated for capillary electrochromatography (CEC). The capillary columns were tested using a commercially available capillary electrophoresis unit. Under typical CEC conditions, the capillary columns packed by gravity showed reduced plate heights below 2 with a reproducibility of less than 9% RSD.

Original languageEnglish
Pages (from-to)1493-1495
Number of pages3
JournalAnalyst
Volume123
Issue number7
DOIs
StatePublished - 1998

Keywords

  • Capillary electrochromatography
  • Gravity-packed columns

Fingerprint

Dive into the research topics of 'Capillary electrochromatography in columns packed by gravity. Preliminary study'. Together they form a unique fingerprint.

Cite this