Abstract
Methods are described for the assessment of patient exposure during clinical fluoroscopic procedures. Values of the roentgen- area-product (RAP) and their distribution throughout the examination are presented for both single-contrast and double-contrast barium enema studies. The double-contrast procedure was measured to give 50% more radiation to the patient than the single contrast procedure when the same size optical aperture is used between the intensifier and TV pick-up tube. However, it was possible to decrease the fluoroscopic RAP value by over a factor of two for the double-contrast procedure without an adverse clinical effect by increasing the area of the aperture diaphragm.
| Original language | English |
|---|---|
| Pages (from-to) | 453-458 |
| Number of pages | 6 |
| Journal | Investigative Radiology |
| Volume | 18 |
| Issue number | 5 |
| DOIs | |
| State | Published - 1983 |