Abstract
A high-performance liquid chromatographic technique primarily developed for use on samples from kidney perfusion studies is presented for simultaneous determination of prednisone, prednisolone and their 20β-hydroxylated metabolites. The technique employs 6β-hydroxycortisol as the internal standard. Samples are extracted with ethyl acetate, washed with sodium hydroxide and water and injected onto a silica gel column with UV detection at 254 nm. Inter- and intraday variability of the assay was determined at two concentrations of each steroid and was less than 10%. Assay steroid recovery ranged from 54.1% for prednisone to 63.2% for 20β-hydroxyprednisone. Sensitivity is 4-10 ng/ml for the steroids measured. The chromatographic conditions may be modified to permit quantitation of these steroids from plasma samples. This method may alternatively be used for quantitation of 6β-hydroxycortisol, an endogenous indicator of enzyme induction. A perfusate concentration-time profile is presented from a kidney perfusion study using prednisolone.