An 8-channel RF coil array for carotid artery MR imaging in humans at 3 T

  • Xiaoqing Hu
  • , Lei Zhang
  • , Xiaoliang Zhang
  • , Huabin Zhu
  • , Xiao Chen
  • , Yongqin Zhang
  • , Yiu Cho Chung
  • , Xin Liu
  • , Hairong Zheng
  • , Ye Li
  • Shenzhen Institute of Advanced Technology
  • Shenzhen Key Laboratory for MRI
  • Ltd.
  • Northwest University China

Research output: Contribution to journal › Article › peer-review

12 Scopus citations

Abstract

Purpose: Carotid artery diseases due to plaque buildup at the carotid bifurcation are a leading cause of stroke. Accurate plaque quantification and characterization of plaque composition and morphology by magnetic resonance imaging (MRI) is essential to identifying high-risk patients. Difficulties in detecting plaque, which is physically small, and the unique physiological structure of the carotid artery make use of a radio frequency (RF) coil array with high resolution, large longitudinal coverage, and deep penetration ideal for clinical examinations. The goal of this project was to design and fabricate a sensitive RF coil array with sufficient imaging coverage and signal-to-noise ratio (SNR) for carotid artery imaging at 3 T. Methods: Based on clinical requirements and the anatomical structure of the human carotid artery, an 8-channel carotid coil array was designed and fabricated for 3 T MRI of the carotid artery in humans. The performance of the proposed 8-channel carotid coil array was validated through bench tests and MR imaging experiments on a 3 T whole body MRI scanner. Its performance was also compared experimentally to the performance of a commercial 4-channel phased array carotid coil designed by Machnet BV (Machnet BV coil, Roden, Netherlands). Results: The 8-channel carotid coil array performed significantly better in imaging the carotid artery than the commercial 4-channel Machnet BV coil in terms of the SNR, coverage, and penetration depth. In parallel imaging, the proposed 8-channel carotid coil array demonstrated a much lower maximum value and average value of the geometry factor in the region of interest. Carotid artery images acquired in vivo using the proposed 8-channel carotid artery coil and the commercial 4-channel Machnet BV coil were also compared, demonstrating the former's potential for clinical diagnosis. Conclusions: Based on the analyses of phantom and in vivo imaging studies, the proposed 8-channel carotid coil array has the potential for use in clinical diagnosis, performing better in terms of SNR, imaging coverage, and penetration depth than the commercial 4-channel carotid artery coil array at 3 T. In future studies, the proposed 8-channel carotid coil array can also serve as an important part of a large-scale multichannel coil array for imaging the whole carotid artery system, including the extracranial and intracranial arteries.

Original languageEnglish
Pages (from-to)1897-1906
Number of pages10
JournalMedical Physics
Volume43
Issue number4
DOIs
StatePublished - 2016

Keywords

  • Carotid artery imaging
  • Magnetic resonance image
  • Multichannel phased array
  • Signal-tonoise ratio