Abstract
Image intensifier distortion due to photocathode curvature and electron optics is shown to be approximated by a simple two parameter odd-power polynomial. The accuracy of this fit was found to be far better than that of two other one parameter characterizations of distortion when applied to experimental data from four different model image intensifiers ranging in diameter from 9 to 14 in. The standard errors of the two parameter fits were less than 0.1 mm or 0.03% of the field of the IIs and were within the estimated measurement error.
| Original language | English |
|---|---|
| Pages (from-to) | 1145-1151 |
| Number of pages | 7 |
| Journal | Medical Physics |
| Volume | 18 |
| Issue number | 6 |
| DOIs | |
| State | Published - Nov 1991 |