Skip to main navigation Skip to search Skip to main content

A programmable pressure control system for coronary flow studies

  • SUNY Buffalo

Research output: Contribution to journal › Article › peer-review

5 Scopus citations

Abstract

An electrohydraulic servo valve capable of reproducing phasic physiological pressure and flow waveforms has been designed and constructed. The device has been utilized successfully in canine coronary flow studies. By measuring coronary pressure directly it is possible to reproduce waveform frequency components as high as 50 Hz in the cannulated left circumflex coronary artery when operated in a negative feedback configuration. It may easily be adapted to operate in a pressure- or flow-controlled mode in any vascular bed. The closed-loop frequency response of the servo valve itself is flat to 100 Hz with a natural frequency of 170 Hz. When utilized in the cannulated bed the frequency response is dependent on perfusion circuit and catheter dynamics.

Original languageEnglish
Pages (from-to)H796-H802
JournalAmerican Journal of Physiology - Heart and Circulatory Physiology
Volume12
Issue number5
DOIs
StatePublished - 1982

Fingerprint

Dive into the research topics of 'A programmable pressure control system for coronary flow studies'. Together they form a unique fingerprint.

Cite this