Skip to main navigation Skip to search Skip to main content

A practical exposure-equivalent metric for instrumentation noise in x-ray imaging systems

  • SUNY Buffalo

Research output: Contribution to journalArticlepeer-review

29 Scopus citations

Abstract

The performance of high-sensitivity x-ray imagers may be limited by additive instrumentation noise rather than by quantum noise when operated at the low exposure rates used in fluoroscopic procedures. The equipment-invasive instrumentation noise measures (in terms of electrons) are generally difficult to make and are potentially not as helpful in clinical practice as would be a direct radiological representation of such noise that may be determined in the field. In this work, we define a clinically relevant representation for instrumentation noise in terms of noise-equivalent detector entrance exposure, termed the instrumentation noise-equivalent exposure (INEE), which can be determined through experimental measurements of noise-variance or signal-to-noise ratio (SNR). The INEE was measured for various detectors, thus demonstrating its usefulness in terms of providing information about the effective operating range of the various detectors. A simulation study is presented to demonstrate the robustness of this metric against post-processing, and its dependence on inherent detector blur. These studies suggest that the INEE may be a practical gauge to determine and compare the range of quantum-limited performance for clinical x-ray detectors of different design, with the implication that detector performance at exposures below the INEE will be instrumentation-noise limited rather than quantum-noise limited.

Original languageEnglish
Pages (from-to)5107-5121
Number of pages15
JournalPhysics in Medicine and Biology
Volume53
Issue number18
DOIs
StatePublished - Sep 21 2008

Fingerprint

Dive into the research topics of 'A practical exposure-equivalent metric for instrumentation noise in x-ray imaging systems'. Together they form a unique fingerprint.

Cite this