A New Algorithm for Minimizing Pacemaker Polarization Artifact: Universally Applicable in Permanent Pacing Systems

  • ANNE B. CURTIS
  • , FRED VANCE
  • , STEPHEN M. QUIST
  • , ALEXANDER DOMIJAN
  • , STEPHEN G. KEIM
  • , AURELIO DURAN
  • , KIM MILLER
  • St. Jude Medical, Inc. Sylmar
  • University of Florida

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

Polarization artifacts that result from pacing may interfere with analysis of paced evoked responses during, e.g., automatic threshold tracking. We have developed a method for reduction of such artifacts that relies on the introduction of pacing stimuli during the refractory period of unipolar or bipolar paced captured beats after previous identification of a refractory period “template” or baseline. The refractory pacing stimuli cannot capture the heart, and thus any deviation from the template is due to polarization artifact alone. The artifact amplitude is measured and the precharge duration of the triphasic stimulus waveform is changed each time until artifact is minimized, as detected by repeated reversals in the polarily of the polarization artifact. In a series of 11 patients with unipolar and bipolar permanent pacing leads, mean initial artifact before balancing was 1.44 ± 0.84 mV, which was reduced to 0.44 ± 0.30 mV after balancing (P = 0.001). Initial precharge duration was 3.2 msec by design; mean final precharge duration was 3.30 ± 0,34 msec. This algorithm is universally applicable in permanent pacing systems, as it is valid in unipolar and bipolar pacing and it does not require an intrinsic cardiac rhythm.

Original languageEnglish
Pages (from-to)1803-1808
Number of pages6
JournalPACE - Pacing and Clinical Electrophysiology
Volume14
Issue number11
DOIs
StatePublished - Nov 1991

Keywords

  • pacemaker stimulus
  • pacemakers
  • polarization artifact

Fingerprint

Dive into the research topics of 'A New Algorithm for Minimizing Pacemaker Polarization Artifact: Universally Applicable in Permanent Pacing Systems'. Together they form a unique fingerprint.

Cite this