Abstract
Due to a lack of fetal imaging coils, the standard commercial abdominal coil is often used for fetal imaging, the performance of which is limited by its insufficient coverage, element number, and Signal-to-noise ratio (SNR). In this paper, a dedicated 36-channel coil array, of which size can best fit the body sizes of pregnancy gestation from 20 to 37+ weeks, was designed for fetal imaging at 3T. SNR with full phase encoding and G-factor denoted as noise amplification for parallel imaging were quantitatively evaluated by phantom studies. Compared with a commercial abdominal coil array, the proposed 36-channel fetal array provides not only SNR improvements in full phase encoding (with 10% in the region where the whole fetal body was located, and up to 40% in the edge region where the fetal brain and heart may appear) but also an augmented parallel imaging capability and remarkable SNR improvements at high acceleration factors.
| Original language | English |
|---|---|
| Article number | 8361840 |
| Pages (from-to) | 2290-2297 |
| Number of pages | 8 |
| Journal | IEEE Transactions on Medical Imaging |
| Volume | 37 |
| Issue number | 10 |
| DOIs | |
| State | Published - Oct 2018 |
Keywords
- Fetus imaging
- magnetic resonance imaging (MRI)
- parallel imaging
- phase array
- radio-frequency (RF) coil