5T magnetic resonance imaging: radio frequency hardware and initial brain imaging

  • Zidong Wei
  • , Qiaoyan Chen
  • , Shihong Han
  • , Shuheng Zhang
  • , Na Zhang
  • , Lei Zhang
  • , Haining Wang
  • , Qiang He
  • , Peng Cao
  • , Xiaoliang Zhang
  • , Dong Liang
  • , Xin Liu
  • , Ye Li
  • , Hairong Zheng
  • Shenzhen Institute of Advanced Technology
  • The Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province
  • University of Chinese Academy of Sciences
  • Shanghai United Imaging Healthcare Co., Ltd.
  • The University of Hong Kong

Research output: Contribution to journalArticlepeer-review

43 Scopus citations

Abstract

Background: We aimed to demonstrate the feasibility of generating high-resolution human brain magnetic resonance imaging (MRI) at 5 Tesla (T) using a quadrature birdcage transmit/48-channel receiver coil assembly. Methods: A quadrature birdcage transmit/48-channel receiver coil assembly was designed for human brain imaging at 5T. The radio frequency (RF) coil assembly was validated by electromagnetic (EM) simulations and phantom imaging experimental studies. The simulated B1+ field inside a human head phantom and inside a human head model generated by the birdcage coils driven in circularly polarized (CP) mode at 3T, 5T and 7T was compared. Signal-to-noise ratio (SNR) maps, the inverse g-factor maps for evaluation of parallel imaging performance, anatomic images, angiography images, vessel wall images and susceptibility weighted images (SWI) were acquired using the RF coil assembly at 5T and compared to those acquired using a 32-channel head coil on a 3T MRI scanner. Results: For the EM simulations, 5T MRI provided less RF inhomogeneity compared to that of 7T. In the phantom imaging study, the distributions of the measured B1+ field were consistent with the distributions of the simulated B1+ field. In the human brain imaging study, the average SNR value of the brain in the transversal plane at 5T was 1.6 times of that at 3T. The 48-channel head coil at 5T had higher parallel acceleration capability than the 32-channel head coil at 3T. The anatomic images at 5T also showed higher SNR than those at 3T. Improved delineation of the hippocampus, lenticulostriate arteries, and basilar arteries was observed at 5T compared to 3T. SWI with a higher resolution of 0.3 mm ×0.3 mm ×1.2 mm could be acquired at 5T, which enabled better visualization of small blood vessels compared to that at 3T. Conclusions: 5T MRI can provide significant SNR improvement compared to that of 3T with less RF inhomogeneity than that of 7T. The ability to obtain high quality in vivo human brain images at 5T using the quadrature birdcage transmit/48-channel receiver coil assembly has significant in clinical and scientific research applications.

Original languageEnglish
Pages (from-to)3232-3240
Number of pages9
JournalQuantitative Imaging in Medicine and Surgery
Volume13
Issue number5
DOIs
StatePublished - May 2023

Keywords

  • 5T
  • Brain imaging
  • high quality imaging
  • magnetic resonance imaging
  • ultrahigh field