Abstract

Purpose: 3D printed patient specific vascular models provide the ability to perform precise and repeatable benchtop experiments with simulated physiological blood flow conditions. This approach can be applied to CT-derived patient geometries to determine coronary flow related parameters such as Fractional Flow Reserve (FFR). To demonstrate the utility of this approach we compared bench-top results with non-invasive CT-derived FFR software based on a computational fluid dynamics algorithm and catheter based FFR measurements. Materials and Methods: Twelve patients for whom catheter angiography was clinically indicated signed written informed consent to CT Angiography (CTA) before their standard care that included coronary angiography (ICA) and conventional FFR (Angio-FFR). The research CTA was used first to determine CT-derived FFR (Vital Images) and second to generate patient specific 3D printed models of the aortic root and three main coronary arteries that were connected to a programmable pulsatile pump. Benchtop FFR was derived from pressures measured proximal and distal to coronary stenosis using pressure transducers. Results: All 12 patients completed the clinical study without any complication, and the three FFR techniques (Angio-FFR, CT-FFR, and Benchtop FFR) are reported for one or two main coronary arteries. The Pearson correlation among Benchtop FFR/Angio-FFR, CT-FFR/Benchtop FFR, and CT-FFR/Angio-FFR are 0.871, 0.877, and 0.927 respectively. Conclusions: 3D printed patient specific cardiovascular models successfully simulated hyperemic blood flow conditions, matching invasive Angio-FFR measurements. This benchtop flow system could be used to validate CTderived FFR diagnostic software, alleviating both cost and risk during invasive procedures.

Original languageEnglish
Title of host publicationMedical Imaging 2018
Subtitle of host publicationBiomedical Applications in Molecular, Structural, and Functional Imaging
EditorsBarjor Gimi, Andrzej Krol
PublisherSPIE
ISBN (Electronic)9781510616455
DOIs
StatePublished - 2018
EventMedical Imaging 2018: Biomedical Applications in Molecular, Structural, and Functional Imaging - Houston, United States
Duration: Feb 11 2018Feb 13 2018

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume10578
ISSN (Print)1605-7422

Conference

ConferenceMedical Imaging 2018: Biomedical Applications in Molecular, Structural, and Functional Imaging
Country/TerritoryUnited States
CityHouston
Period02/11/1802/13/18

Fingerprint

Dive into the research topics of '3D printed cardiovascular patient specific phantoms used for clinical validation of a CT-derived FFR diagnostic software'. Together they form a unique fingerprint.

Cite this